Author: Ying Zhu (United States)
Co-authors: Vara Wuyyuru, Val Kolesnitchenko
Purpose
To evaluate intraocular pressure (IOP) control performance of 27 Gauge (Ga) dual-cutting, 20K cuts per minute (cpm) beveled vitrectomy probes under different system settings; to compare IOP control performance for 27 Ga dual-cutting, 20K cpm beveled vitrectomy probes with previous generation 27 Ga single-cutting, 10K cpm beveled vitrectomy probes
Setting/Venue
Various laboratory bench tests
Methods
27 Ga dual-cutting 20K cpm beveled and single-cutting 10K cpm beveled vitrectomy probes were driven by dual-pneumatic vitrectomy system with IOP control to aspirate sterile irrigating solution in a hollow acrylic eye model. A digital transducer (OMEGA, PX409-001GUSBH) was connected to the bottom of eye model to detect IOP change during aspiration. Six samples were tested under core duty cycle and vacuums of 250 mmHg, 450 mmHg and 650 mmHg. Cut rates ranged from 2,500 cpm to 10,000 cpm for 10K probes and from 2,500 cpm to 20,000 cpm for 20K vitrectors. Both system IOP compensation enabled and disabled settings were used. Average IOP during aspiration was calculated for each test setting and statistical analyses were performed using Welch’s t-test with statistical significance level of p<0.05
Results
Without IOP compensation, IOP for 27 Ga 20K probes was similar for all cut rates. IOP ranged from 22.71 ± 0.30 mmHg to 22.81 ± 0.37 mmHg for 250 mmHg, 15.46 ± 0.33 mmHg to 15.62 ± 0.33 mmHg for 450 mmHg and 7.93 ± 0.46 mmHg to 8.33 ± 0.32 mmHg for 650 mmHg. However, IOP for 10K probes was significantly different when cut rates changed (p<0.05). IOP ranged from 25.47 ± 0.38 mmHg to 27.46 ± 0.43 mmHg for 250 mmHg, 20.51 ± 0.54 mmHg to 24.32 ± 0.73 mmHg for 450 mmHg, and 16.14 ± 0.77 mmHg to 19.30 ± 0.77 mmHg for 650 mmHg. When IOP control was enabled, there was no significant difference of IOP when using different vacuum and cut rate for 20K probes and 10K probes (p>0.05). IOP for 20K probes at maximum cut rate significantly increased to 29.24 ± 0.75 mmHg for 250 mmHg, 28.43 ± 1.71 mmHg for 450 mmHg and 27.42 ± 2.64 mmHg for 650 mmHg as compared with results without IOP compensation (p<0.05). IOP of 10K probes was similar to that of 20K probes with IOP compensation.
Conlusions
27 Ga dual-cutting 20K cpm vitrectomy probes provide a more constant IOP level compared to the previous generation single-cutting 10K cpm vitrectors under different cut rates without IOP compensation. Using IOP compensation and 27 Ga dual-cutting 20K cpm vitrectomy probes can help surgeons keep the eye at stabilized IOP ranges and have an efficient aspiration.
Financial Disclosure
This is Alcon funded study. Ying Zhu, Val Kolesnitchenko and Vara Wuyyuru are employees of Alcon