Defining appropriate parameters for robotic subretinal drug delivery ensures minimal reflux
Author: Reza Ladha (Belgium)
Co-authors: Jorrit Smit, Esmée Cerfonteijn, Thijs Meenink, Marc D de Smet
Purpose
Vitreous reflux during subretinal injection limits the degree and surface area transfected and is wasteful of a precious product. The precision and stability offered by the Preceyes platform allows for an optimized, time-independent delivery. The purpose of this study is to identify parameters important to insure reflux-free robotic subretinal drug delivery.
Setting/Venue
Comparative trial in an ex vivo animal model
Methods
Freshly harvested porcine eyes underwent a standard pars plana vitrectomy with IOP control under microscopic visualization through a Rescan 700 Zeiss device; followed by subretinal injection using a 41G Teflon needle, with the Preceyes Surgical System (PSS). The following subretinal injection cannula parameters were analyzed: speed of injection, tip configuration, and positioning.
Results
Following multiple injections, positioning of the tip above the RPE without penetration through Bruch’s membrane was possible in all cases by using a stepwise penetration algorithm, irrespective of the angle of approach. Blunt needles did not penetrate as well as needle tips with polished edges. Small robotically controlled rotations and wiggling of the injection cannula did not result in a cleaner penetration or less reflux. Sufficient delay is required between the completion of the injection and retraction of the needle to prevent reflux.
Conlusions
Robotics allows for a better understanding of the injection process and a determination of critical parameters and steps. Several parameters appear critical to an efficient reflux-free injection.
Financial Disclosure
Preceyes BV - support + access to the technology for research purposes