Author: José Ignacio Fernandez-Vigo (Spain)
Co-authors: Francisco Javier Moreno-Morillo, Emilio López-Varela, Jorge Novo-Bujan, Marcos Ortega-Hortas, Lorenzo López-Guajardo, Juan Donate-López
Purpose
To assess the repeatability of the flow signal voids (FSV) measurements of the choriocapillaris (CC) and choroid (CH) in patients with central serous chorioretinopathy (CSCR) by Swept-Source optical coherence tomography angiography (SS-OCTA).
Setting/Venue
Hospital Clínico San Carlos, Madrid, Spain and VARPA Research Group, Instituto de Investigación Biomédica de A Coruña (INIBIC), Universidade da Coruña, A Coruña, Spain.
Methods
Cross-sectional study including 52 eyes with unilateral CSCR. Both eyes of each patient were included to assess the reproducibility of the FSV analysis in the affected eyes with persistent subretinal fluid (SRF) (CSCR group) and the fellow unaffected eyes (non-CSCR group). Two consecutive macular scans in the same eye with an interval of two minutes using the 6 x6 mm scan of a SS-OCTA device (Plex Elite 9000; Zeiss, Dublin, CA) were obtained. FSV area and the number of contours measurements were analyzed using a customized software in three slabs: upper CC (4 to 20 µm below the retinal pigment epithelium, RPE), lower CC (29 to 49 µm below the RPE) and choroid (64 to 115 µm below the RPE). The repeatability of the measurements was assessed as intraclass correlation coefficients (ICC) and coefficients of variation (CV).
Results
Mean FSV area in the CSCR eyes was 5.319 ±0.934 µm2 (3.041 – 7.645) and 5.316 ±0.942 µm2 (3.473 – 8.008) for the upper CC, 1.998 ±0.557 (1.136 - 3.899) and 1.998 ±0.576 µm2 (1.095 - 4.018) for the lower CC and 0.538 ±0.349 µm2 (0.193 - 2.056) and 0.531 ±0.333 µm2 (0.157 - 1.850) for the CH for the measurement 1 and 2 respectively. In the CSCR group, ICCs for the FSV area in the three slabs were all ≥0.858. Higher values were obtained for the lower CC and the CH (ICC= 0.959 and 0.964 respectively) than for the upper CC (0.859). Similar ICC values were obtained for the non-CSCR eyes, observing the highest values for the lower CC 0.949 (0.913 - 0.970), followed by the CH 0.932 (0.885 - 0.961) and upper CC 0.844 (0.744 - 0.908). Regarding the FSV number of contours measurements, the ICCs were all ≥0.844. Higher values were obtained for lower CC and CH (ICC= 0.949 and 0.932 respectively) than for the upper CC (0.844). CV was 4.7%, 3.8% and 8.6% in the CSCR eyes and 4.8%, 3.9% and 10.3% in the non-CSCR group for the upper CC, lower CC and CH respectively.
Conlusions
SS-OCTA offers good repeatability to quantify macular FSV, both in CSCR eyes, even with the presence of SRF, and fellow eyes. Both segmentation slabs for the CC showed good reproducibility and low coefficient of variation, having also good repeatability but higher coefficient of variation the choroidal FSV quantification. Therefore, this instrument could be useful to reliably assess changes in the perfusion of the CC and CH during follow-up or after different treatment approaches.