That really flows?! – A closer look at the cutting process of a newly developed laser vitrectomy system
Author: Daniel Uthoff (Germany)
Co-authors: Rudolf Walker, Norbert Schrage, Jörg Meister
Purpose
The movement of the duty-cycle of existing mechanical vitrectomy cutters causes traction through the vitreous body. This force is transmitted to the retina and can lead to tears. Replacing this mechanical movement with a laser-induced plasma to cut the vitreous, these tractions could be minimised and allow continuous flow at the vitrectomy tip.
Setting/Venue
ultraslowmotion videos show the dutycycle of a mechanical and laser vitrectomy-cutter removing porcine vitreous and visualized balanced salty solution (BSS). Laser source is a diode-pumped Q-switched Nd:YAG laser, λ = 1064 nm, pulse duration 4 ns.
Methods
The dutycycle and vitreous removal are described with video recordings and physical calculations; what is the duration of the cutter opening and cutter closure, how exactly do the cutting process and the flow interact?
Results
Video recordings show an almost continuous flow of tissue at the laser-vitrectomy-tip and present the differences to mechanical cutting systems.
Conlusions
The use of a laser cutter allows less traction within the vitreous body, a reduction in iatrogenic complications during vitrectomy is therefore conceivable.