The use of microperimetry in assessing visual function after vitrectomy with endotamponade due to reghmatogenous retinal detachment (RRD).
Author: Anna Sukhanova (Russian Federation)
Co-authors: Rinat Fayzrakhmanov
Purpose
To evaluate changes in retinal sensitivity after vitrectomy with gas and silicone oil tamponade due to rhegmatogenous retinal detachment.
Setting/Venue
N. I. Pirogov National Medical and Surgical Center of the Ministry of Healthcare of the Russian Federation.
Methods
20 cases of surgery treatment of patients with primary macula-off RRD were enrolled. All cases were divided in two groups depending on the type of endotamponade. In the 1st group (10 eyes) a vitreoretinal surgery with silicone oil tamponade (1300 cSt) was performed. The 2d group (10 eyes) included only cases with a gas (C2F6) tamponade. The control group included contralateral eyes without ophthalmic pathology. An outcome was best-corrected visual acuity (BCVA) and retinal sensitivity. The examination was performed 30 days after silicone oil removal for the 1st group and 30 days after a primary surgery with a gas tamponade for the 2d group. The participants underwent microperimetry in the expert mode 10°. The mean sensitivity threshold was calculated for each ring (zone I-IV) of the scanning grid, where zone I corresponded to 0°.
Results
The duration of the detachment did not differ between the groups (р =0,23). The mean macular sensitivity threshold in the 1st group was significantly reduced (19,7 ± 2,7* dB), both in comparison with the 2d group (p = 0.007) and the control group (p = 0.003). Differentiation by zones in the 1st group revealed a decrease in each analyzed zone in comparison with the control group and a decrease in zone II (p = 0.031) and in zone IV (p = 0.038) in comparison with the 2d group. In zone I of the 1st group the relative scotoma was revealed in 5 cases out of 15 (33%). The mean macular sensitivity threshold in the 2d group was significantly reduced in each zone in comparison with the control group. A strong direct correlation was found between BCVA and mean macular sensitivity threshold in the 2d group (r = 0.87).
Conlusions
There is a decrease in the functional parameters of the retina after vitrectomy with silicone oil in comparison with gas tamponade. This decrease is in the form of a reduction of retinal sensitivity in the 10° zone according to microperimetry data. Microperimetry is the preferred research method when assessing the dynamics of the functional parameters of the retina after vitrectomy with various types of tamponade.